Laboratory & Analytical

Laboratory Water Purification Systems — ASTM D1193 Type I–III & ISO 3696 Grade 1–3, 10 L/h to 2,000 L/h

Point-of-use purifiers and central laboratory loops engineered to the grade your instruments actually need — Type I for HPLC, ICP-MS and cell culture, Type II for reagent make-up and microbiology, Type III for glassware washers and autoclave feed. RO → EDI → 254 nm UV → 0.22 µm point-of-use polish, with terminal ultrafiltration where endotoxin control is required.

Request a Lab Water Grade Map Talk to a lab water specialist
18.2 MΩ·cm Type ITOC < 10 ppbBacteria < 1 CFU/mLEndotoxin UF on request
laboratory application — water treatment equipment detail
Process water system — key stage
laboratory application — process component
System component in operation
laboratory application — process component
Process detail view
laboratory client / installation site visit
On-site / client visit
>

Why Lab Water Is Bought by Grade, Not by Flow Rate

Almost every lab water quote we are asked to beat is compared on litres per hour. That is the wrong axis. An ICP-MS will not care that your purifier makes 60 L/h if the TOC is drifting; a cell culture lab will lose a week of work to bacteria long before it runs out of water. The right question is which grade each instrument needs, then how much peak flow those instruments draw together. We map that before quoting a single skid.

GradeTypical useTarget we hold
Type I (ASTM D1193)HPLC, ICP-MS / ICP-OES, GC-MS, cell culture, molecular biology18.2 MΩ·cm @ 25 °C, TOC < 10 ppb, bacteria < 1 CFU/mL
Type IIReagent make-up, microbiology media, general chemistry, feed to Type I> 1 MΩ·cm, TOC < 50 ppb, bacteria < 100 CFU/mL
Type IIIGlassware rinsing, autoclave / washer feed, humidifier feed> 4 MΩ·cm, TOC < 200 ppb
CLRW (clinical, CLSI GP40)Clinical chemistry, immunoassay analysers≥ 10 MΩ·cm, bacteria < 10 CFU/mL, endotoxin controlled by terminal UF
Endotoxin-sensitiveCell culture, pyrogen-sensitive workTerminal 10 kDa UF; validated against your lab's own limit
Peak flowGlassware washer + several taps simultaneouslySized on simultaneous draw, not average

Our Standard Purification Train for Laboratories

StageEquipmentFunctionKey spec
① Pretreatment5 µm sediment + activated carbon + softener (or antiscalant)Protect RO from chlorine, hardness, particlesFree Cl₂ < 0.1 ppm, SDI < 5
② ROCompact RO module (e.g. RO-250 class)Bulk demineralisation, 95–99% rejectionPermeate < 20 µS/cm
③ DeionisationEDI or exchangeable DI cylindersPolish to > 15 MΩ·cmEDI: continuous; cylinders: scheduled exchange
④ Bacteria control254 nm UV + recirculationSuppress biofilm in storage and loopDose ≥ 30 mJ/cm²
⑤ TOC reduction185 nm UV (on Type I systems)Oxidise trace organicsTOC < 10 ppb at point-of-use
⑥ PolishingNuclear-grade mixed-bed cartridgeFinal ionic polish to 18.2 MΩ·cmReplace on resistivity or timer
⑦ Terminal filtration0.22 µm capsule / 10 kDa UFBacteria and endotoxin removal at the tapBacteria < 1 CFU/mL
⑧ DistributionPVDF or electropolished SS316L loopDeliver without recontamination1.5–2 m/s, zero dead legs
⑨ SanitisationChemical or hot-water / ozone cyclePeriodic biofilm controlQuarterly or per your SOP
⑩ MonitoringOnline resistivity + TOC + flowEvidence for accreditationData log exportable to LIMS

Small labs with a single instrument room usually stop at ⑦ with a point-of-use unit. Multi-floor QC labs and hospitals need ⑧–⑩, and that is where most "cheap" systems fail their first audit.

Materials We Use (Leachables Matter Here)

Wetted partMaterialWhy it is chosen
Distribution loopPVDF or electropolished SS316LLowest extractables, sanitisation-tolerant
Final run to the tapPFA or PVDFNo plasticiser leaching into Type I water
Storage tankHDPE or PVDF-lined, vent-filteredNo metal contact, bacteria-tight vent
ValvesDiaphragm (PTFE / EPDM) or PVDFNo dead legs, serviceable in place
Point-of-use housingMedical-grade PP / PVDFClean replacement without tools
UF module10 kDa polysulfone / PES hollow fibreEndotoxin and nuclease retention

Material and extractables documentation ships with every system — useful evidence when your accreditation body asks why the loop is PVDF and not PVC.

Standard Capacities (Laboratory & Analytical)

ModelOutputPowerTypical application
LB-1010 L/h0.3 kWSingle instrument bench, HPLC
LB-3030 L/h0.5 kWSmall QC room, 3–5 benches
LB-6060 L/h0.9 kWAnalytical lab + glassware washer
LB-120120 L/h1.6 kWOne-floor loop, mixed Type I / II draws
LB-250250 L/h3.2 kWMulti-floor QC or hospital lab
LB-500 / LB-1000+500–2,000 L/h6–18 kWCentral ring main, campus or hospital

Point-of-use units ship pre-assembled and are typically running the same day. Central loops include loop design, installation and a documented sanitisation and monitoring SOP.

Engineering Reference — 120 L/h Loop for a Three-Floor QC Laboratory

Engineering reference design for this capacity class. Actual delivered laboratory projects are summarised on our About page under mutual NDA; full documentation available on request.
ParameterDesign value
FeedCity water, TDS ≈ 220 ppm, free Cl₂ < 0.5 ppm, hardness ≈ 2.1 mmol/L
Type I demand40 L/h continuous, 2× peak for 15 min
Type II demand60 L/h (media prep, reagent make-up)
Type III demand20 L/h (glassware washer, autoclave feed)
Product quality18.2 MΩ·cm at point-of-use, TOC < 10 ppb, bacteria < 1 CFU/mL
LoopPVDF ring main, 1.8 m/s, 3× tank turnover/h, zero dead legs
Storage300 L PVDF-lined tank, vent filter, 254 nm UV on return
MonitoringOnline resistivity + TOC, logged to LIMS over Modbus TCP
SanitisationChemical cycle quarterly, ozone on request
Delivery6–8 weeks from PO

5 Pitfalls We Design Around

  • Buying on 18.2 MΩ·cm alone — resistivity says nothing about TOC, bacteria or nucleases. We specify all four together and log the two that drift (TOC and bacteria).
  • Bacteria rebound in the storage tank — a tank without UV on the return and a vent filter becomes a biofilm reactor. We size 254 nm UV on the return and specify the sanitisation interval in the SOP.
  • PVC or wrong-plastic extractables — cheap tubing leaches plasticisers into Type I water and shows up as ghost peaks. We use PVDF or electropolished SS316L on the loop and PFA on the final drop.
  • Undersized peak draw — a glassware washer pulling 20 L/min collapses loop pressure and everyone's resistivity dips. We size on simultaneous demand, not on average consumption.
  • No sanitisation routine — every loop needs a documented cycle (chemical, hot water or ozone) and a replacement schedule for final filters. We hand over the schedule, not just the machine.

Standards We Design To

  • ASTM D1193 — Standard specification for reagent water (Type I–IV)
  • ISO 3696 — Water for analytical laboratory use (Grade 1–3)
  • CLSI GP40 — Preparation and testing of reagent water in the clinical laboratory (CLRW)
  • CAP / ISO 15189 checklist requirements for clinical laboratory water (documentation support)
  • ISO 9001 / ISO 14001 / ISO 45001 certified factory; CE / IEC 61010 for instrumentation
  • EN 10204 3.1 material certificates on loop and tank wetted parts

Frequently Asked Questions

Q1. Which grade does each of our instruments need?

Rule of thumb: Type I for HPLC, ICP-MS, GC-MS, cell and molecular biology; Type II for reagent make-up, media and microbiology; Type III for glassware washers and autoclaves. Send us your instrument list and we will return a grade-by-grade map with the peak flow for each.

Q2. EDI or replaceable DI cylinders?

Cylinders are cheaper to buy and fine for a single bench that uses under about 30 L/h. EDI is continuous, needs no cylinder exchange and no waste resin handling, and is the better choice once you cross roughly 60 L/h or run a loop.

Q3. Do we need a distribution loop, or will point-of-use units do?

Point-of-use units win below about 60 L/h and when instruments sit in one room. Once draws are spread over more than one floor, or you need documented monitoring for accreditation, a PVDF loop with online resistivity and TOC is the defensible design.

Q4. How do you control bacteria long term?

Three things together: 254 nm UV on the tank return, continuous recirculation at 1.5–2 m/s with no dead legs, and a documented sanitisation cycle (chemical, hot water or ozone) plus scheduled final-filter replacement.

Q5. Can you meet endotoxin requirements for cell culture or clinical work?

Yes — terminal 10 kDa ultrafiltration at the point of use removes endotoxin and nucleases. We size the UF module to your validated limit and provide retention data with the unit.

Q6. What monitoring and documentation do we get?

Online resistivity and TOC with data logging, Modbus TCP export to your LIMS, a full O&M manual, EN 10204 3.1 material certificates, and a commissioning report with IQ/OQ templates suitable for ISO 15189 or GMP-adjacent audits.

Q7. What feed water do you accept?

Potable city water, TDS up to about 500 ppm, free chlorine under 1 ppm. Hard water gets a softener or antiscalant ahead of the RO; very hard or high-silica well water may need pretreatment designed case by case.

Q8. What is the warranty and consumables position?

12 months on the complete system plus 36 months on critical parts (RO membrane, EDI stack, UV lamps). We ship a 2-year consumables forecast with every system, so filter and lamp changes are budgeted rather than emergency purchases.

Ready to Specify Your Lab Water Grades?

Send your instrument list — we return a free grade map (which bench needs Type I, II or III) with peak-flow sizing within 24 h.

Request a Lab Water Grade Map Talk to a lab water specialist

Technical content reviewed by Ella Liu, Industrial Water Treatment Specialist at Rong Water — 10+ years specifying Type I–III and clinical reagent water systems for QC, analytical and hospital laboratories.

Get a Quote
Scroll to Top